Functional Checks To Be Performed After Replacement Procedures; General Overview; Submitting A Replacement Procedure Report; Functional Checks Required Per Replacement Part Category - GE Venue Go Service Manual

Table of Contents

Advertisement

P R E L I M I N A R Y
D
5813707-100, R
IRECTION
Section 8-4

Functional Checks to be Performed after Replacement Procedures

8-4-1

General Overview

Table 8-2
For easy reference, these are grouped in accordance with the various categories of Venue Go™
replacement parts.
IMPORTANT - AFTER PERFORMING A REPLACEMENT PROCEDURE, ALWAYS REPORT TO
NOTICE
THE GE SERVICE SYSTEM, AS DESCRIBED BELOW.
8-4-2

Submitting a Replacement Procedure Report

After performing a Venue Go™ Replacement Procedure, if all required Functional Checks passed
successfully, proceed as follows to submit a Debrief Script to the GE Service System:
Using your pole's Dispatch Tool, enter the following script in the Service Comments when debriefing
a Service Dispatch:
Venue Go™ Service Manual, Direction Direction 5813707-100, Revision 2 - per
82. Equipment passed all required tests and is ready for use.
8-4-3

Functional Checks Required per Replacement Part Category

Table 8-2
Part Number
System Parts
S5806961
S5764348
S5775296
S5775292
S5762368
S5809900
S5789349
S5797835
8-82
2
EVISION
below lists the Functional Checks to be performed after each Replacement Procedure.
lists the Functional Checks to be performed after each type of Replacement Procedure.
Replacement Element and Procedure
PROBE HOLDER RUBBER - Venue Go
Batteries
Back cover FRU - Venue Go
Speakers
C-PSB FRU - Venue Go
RS Probe Latches FRU - Venue Go
Venue Go Fan FRU - Venue Go
WiFi PCB FRU - Venue Go
Table 8-2 Replacement Procedures - Functional Checks Required
Chapter 8 - Replacement Procedures
V
Functional Checks Required
• Visual inspection
4-3-9-1 -Battery Charging Test
7-5-3 -Running System Diagnostics
4-3-5 -Image Quality Tests
4-3-8 -Mechanical Functions Checks
7-5-3 -Running System Diagnostics
4-3-5 -Image Quality Tests
7-5-3 -Running System Diagnostics
4-3-5 -Image Quality Tests
• Visual inspection
• Visual inspection
7-5-3 -Running System Diagnostics
4-3-5 -Image Quality Tests
• Wifi test
7-5-3 -Running System Diagnostics
4-3-5 -Image Quality Tests
™ S
M
ENUE
ERVICE
ANUAL
Table 8-2
on page 8-

Advertisement