GE HEALTHCARE
REV 8
SECTION 3
FUNCTIONAL CHECK
Inspect
3−1
Power on
ON
OFF
3−2
Drive
3−3
Tube Column and
Arm
Requirements are pass or fail. Service is required if any Requirement is failed. After each
Requirement is verified, place a check in its box. Record notes and comments next to
the Requirement.
Requirement
□
Changing Key Switch position from off to on
makes the Operators Display flash, then display
TESTING COMPLETE.
Recharge if display reads RECHARGE
RECOMMENDED or RECHARGE IMMEDIATELY.
□ Pressing kVp UP increases kVp display.
□ Pressing kVp DOWN decreases kVp display.
□ Pressing mAs UP increases mAs display.
□ Pressing mAs DOWN decreases mAs display.
□
Drive Enable bar raises and lowers easily.
□ Drives forward, reverse, left, and right.
□ Drive speed is easily controlled.
□ Drives only in reverse when bumper is pushed in.
□ Drive speed reduced when tube arm is
removed from latch.
□ Drive brakes release when brake release
switch is depressed.
□
Horizontal Tube Arm Latch operates smoothly
and holds arm securely.
□ Vertical Tube Column rotates 270 degrees from
latch position to latch position.
□ Tube Arm moves smoothly from bottom to top
of tube column.
□ Tube Arm extends and retracts smoothly.
□ Tube Unit rotates 90 degrees to horizontal in
both directions.
□ Tube Unit rotates back 10 and forward 110
degrees.
□ Latch holds column in drive position.
□ Lock prevents vertical travel.
□ Lock prevents tube extension.
□ Lock prevents column rotation.
19
AMX−4+ FUNCTIONAL CHECK
(MODEL 2169360, 2236420 & 2275938 SERIES)
DIRECTION 2173224−100
Inspector's Notes