Section 3 Functional Check; Power On; Drive; Tube Column And Arm - GE AMX-4+ Functional Check Instructions

Hide thumbs Also See for AMX-4+:
Table of Contents

Advertisement

GE HEALTHCARE
REV 8
SECTION 3
FUNCTIONAL CHECK
Inspect
3−1

Power on

ON
OFF
3−2

Drive

3−3
Tube Column and
Arm
Requirements are pass or fail. Service is required if any Requirement is failed. After each
Requirement is verified, place a check in its box. Record notes and comments next to
the Requirement.
Requirement
Changing Key Switch position from off to on
makes the Operators Display flash, then display
TESTING COMPLETE.
Recharge if display reads RECHARGE
RECOMMENDED or RECHARGE IMMEDIATELY.
□ Pressing kVp UP increases kVp display.
□ Pressing kVp DOWN decreases kVp display.
□ Pressing mAs UP increases mAs display.
□ Pressing mAs DOWN decreases mAs display.
Drive Enable bar raises and lowers easily.
□ Drives forward, reverse, left, and right.
□ Drive speed is easily controlled.
□ Drives only in reverse when bumper is pushed in.
□ Drive speed reduced when tube arm is
removed from latch.
□ Drive brakes release when brake release
switch is depressed.
Horizontal Tube Arm Latch operates smoothly
and holds arm securely.
□ Vertical Tube Column rotates 270 degrees from
latch position to latch position.
□ Tube Arm moves smoothly from bottom to top
of tube column.
□ Tube Arm extends and retracts smoothly.
□ Tube Unit rotates 90 degrees to horizontal in
both directions.
□ Tube Unit rotates back 10 and forward 110
degrees.
□ Latch holds column in drive position.
□ Lock prevents vertical travel.
□ Lock prevents tube extension.
□ Lock prevents column rotation.
19
AMX−4+ FUNCTIONAL CHECK
(MODEL 2169360, 2236420 & 2275938 SERIES)
DIRECTION 2173224−100
Inspector's Notes