Dosimass Document information Document information Document function These Operating Instructions contain all the information that is required in various phases of the life cycle of the device: from product identification, incoming acceptance and storage, to mounting, connection, operation and commissioning through to troubleshooting, maintenance and disposal.
• The W@M Device Viewer : Enter the serial number from the nameplate (www.endress.com/deviceviewer) • The Endress+Hauser Operations App: Enter the serial number from the nameplate or scan the 2-D matrix code (QR code) on the nameplate. For a detailed list of the individual documents along with the documentation code...
Registered trademark of the Microsoft Corporation, Redmond, Washington, USA VITON ® Registered trademark of DuPont Performance Elastomers L.L.C., Wilmington, USA TRI-CLAMP ® Registered trademark of Ladish & Co., Inc., Kenosha, USA Applicator , DeviceCare , FieldCare ® ® ® Registered or registration-pending trademarks of the Endress+Hauser Group Endress+Hauser...
‣ Observe the specified pressure and temperature range. Verification for borderline cases: ‣ For special fluids and fluids for cleaning, Endress+Hauser is glad to provide assistance in verifying the corrosion resistance of fluid-wetted materials, but does not accept any Endress+Hauser...
It meets general safety standards and legal requirements. It also complies with the EC directives listed in the device-specific EC Declaration of Conformity. Endress+Hauser confirms this by affixing the CE mark to the device.
Page 10
Basic safety instructions Dosimass IT security measures in line with operators' security standards and designed to provide additional protection for the device and device data transfer must be implemented by the operators themselves. Endress+Hauser...
• If one of the conditions is not satisfied, contact your Endress+Hauser Sales Center. • Depending on the device version, the CD-ROM might not be part of the delivery! In such cases, the technical documentation is available via the Internet or via the Endress+Hauser Operations App, see the "Product identification"...
"Supplementary device-dependent documentation" (→ • The W@M Device Viewer: Enter the serial number from the nameplate (www.endress.com/deviceviewer) • The Endress+Hauser Operations App: Enter the serial number from the nameplate or scan the 2-D matrix code (QR code) on the nameplate. 3.2.1...
Dosimass Incoming acceptance and product identification 3.2.2 Symbols on measuring device Symbol Meaning WARNING! This symbol alerts you to a dangerous situation. Failure to avoid this situation can result in serious or fatal injury. Reference to documentation Refers to the corresponding device documentation.
Storage and transport Dosimass Storage and transport Storage conditions Observe the following notes for storage: • Store in the original packaging to ensure protection from shock. • Do not remove protective covers or protective caps installed on process connections. They prevent mechanical damage to the sealing surfaces and fouling in the measuring tube.
Dosimass Storage and transport 4.2.2 Measuring devices with lifting lugs CAUTION Special transportation instructions for devices with lifting lugs ‣ Only use the lifting lugs fitted on the device or flanges to transport the device. ‣ The device must always be secured at two lifting lugs at least.
Installation Dosimass Installation Installation conditions No special measures such as supports are necessary. External forces are absorbed by the construction of the device. 5.1.1 Mounting position Mounting location To prevent measuring errors arising from accumulation of gas bubbles in the measuring tube, avoid the following mounting locations in the pipe: •...
Page 17
Dosimass Installation Ø orifice plate, pipe restriction [mm] [in] [mm] [in] ³⁄₈ 0.24 ½ 0.40 0.55 Orientation The direction of the arrow on the sensor nameplate helps you to install the sensor according to the flow direction (direction of medium flow through the piping).
Installation Dosimass A0003830 Recommended orientation in the event of extreme heating Inlet and outlet runs No special precautions need to be taken for fittings which create turbulence, such as valves, elbows or T-pieces, as long as no cavitation occurs (→...
Page 19
Dosimass Installation A0015594 Thermal insulation In the case of some fluids, it is important that the heat radiated from the sensor to the transmitter is kept to a minimum. A wide range of materials can be used for the required insulation.
Installation Dosimass 5.1.3 Special mounting instructions Zero point adjustment All measuring devices are calibrated in accordance with state-of-the-art technology. Calibration takes place under reference conditions (→ 57). Therefore, a zero point adjustment in the field is generally not required. Experience shows that zero point adjustment is advisable only in special cases: •...
Dosimass Installation Post-installation check Is the device undamaged (visual inspection)? Does the measuring device conform to the measuring point specifications? For example: • Process temperature (→ • Process pressure (refer to the chapter on "Pressure-temperature ratings" of the "Technical Information"...
Electrical connection Dosimass Electrical connection Connection conditions WARNING Risk of electric shock. ‣ Have electrical connection work carried out by correspondingly trained specialists only. ‣ Observe applicable federal/national installation codes and regulations. ‣ Comply with local workplace safety regulations. ‣...
Dosimass Electrical connection 6.2.1 Pin and socket assignment Pulse/frequency/status output (option 3) A0023690 Connection (option 3) Option 3: Pulse/frequency/status output Assignment RSE8 M12 × 1 L+ Supply voltage: 24 V nominal voltage (20 to 30 V ), 3.5 W RX service interface (may not be connected during normal operation)
Page 24
Electrical connection Dosimass Connection option 3 A0023237 8-pin device connection Socket, input Connector, input Supply voltage + Service interface Service interface Supply voltage – (+) pulse/frequency/status output (–) pulse/frequency/status output (–) pulse/frequency/status output Service interface PELV or SELV power supply...
Dosimass Electrical connection 6.2.3 Cable specifications Use connecting cables with a cross-section of at least 0.25 mm (0.0004 in ) (e.g. AWG23). The temperature specification of the cable must be at least 20 °C (68 °F) higher than the maximum ambient temperature in the application.
• DVD (contact Endress+Hauser) To ensure the correct operation of the DeviceCare and FieldCare programs, the computer running the programs must meet certain minimum requirements with regard to hardware and software. The minimum requirements are listed on the Endress+Hauser website www.endress.com. Endress+Hauser...
If startup fails, an error message to this effect is displayed in the DeviceCare or FieldCare operating program, depending on the cause. 8.2.1 Zero point adjustment All Dosimass measuring devices are calibrated in accordance with state-of-the-art technology. Calibration takes place under reference conditions (→ 57). Dosimass therefore does not require zero point adjustment! Experience shows that zero point adjustment is advisable only in special cases: •...
Commissioning Dosimass A0008558 Performing the zero point adjustment 1. Let the system run until normal operating conditions are present. 2. Stop the flow (v = 0 m/s (0 ft/s)). 3. Check the shutoff valves for leaks. 4. Check the necessary process pressure.
Dosimass Commissioning ‣ Partially filled pipe detection ‣ Advanced setup (→ 8.3.1 Defining the tag name To enable fast identification of the measuring point within the system, you can enter a unique designation using the Device tag parameter and thus change the factory setting.
Commissioning Dosimass Parameter overview with brief description Parameter Description Selection Factory setting Mass flow unit Select mass flow unit. Unit choose list Country-specific: • 75 = kg/h Result • 81 = lb/min The selected unit applies for: • Output • Low flow cut off •...
Page 31
Dosimass Commissioning Switch output function Assign diagnostic behavior Assign limit Assign flow direction check Assign status Value per pulse Pulse width Failure mode Minimum frequency value Maximum frequency value Measuring value at maximum frequency Failure mode Failure frequency Switch-on value...
Commissioning Dosimass Parameter Description Selection / User entry Factory setting Assign frequency output Select process variable for frequency output. • 0 = Off 0 = Off • 1 = Mass flow • 2 = Volume flow • 4 = Density • 7 = Temperature Switch output function Select function for switch output.
Dosimass Commissioning Low flow cut off Assign process variable On value low flow cutoff Off value low flow cutoff Pressure shock suppression Parameter overview with brief description Parameter Description Selection / User entry Factory setting Assign process variable Select process variable for low flow cut off.
Dosimass Commissioning Parameter overview with brief description Parameter Description Selection Factory setting Installation direction Set sign of flow direction to match the • 0 = Flow in arrow direction 0 = Flow in arrow direction direction of the arrow on the sensor.
Commissioning Dosimass Parameter overview with brief description Parameter Description Selection Factory setting Assign process variable Select process variable for totalizer. • 0 = Off 1 = Mass flow • 2 = Volume flow • 1 = Mass flow Mass unit Select mass unit.
Page 37
Dosimass Commissioning Simulation device alarm Simulation diagnostic event Parameter overview with brief description Parameter Prerequsite Description Selection / User Factory setting entry Assign simulation process variable – Select a process variable for • 0 = Off 0 = Off the simulation process that is •...
Operation Dosimass Operation Operation The Operation menu with its submenus contains all the parameters needed for operation. Navigation "Operation" menu Operation Access status tooling (→ Locking status (→ ‣ Totalizer handling (→ ‣ Batching 9.1.1 Operating software access authorization The following access authorization can be configured with the Access status tooling parameter.
Dosimass Operation Totalizer operation The Totalizer handling submenu contains all the parameters that are needed for the totalizer. 9.2.1 Control totalizer The Totalizer handling submenu contains all the parameters that are needed for totalizer control. Navigation "Operation" menu → Totalizer handling...
Diagnostics and troubleshooting Dosimass Diagnostics and troubleshooting 10.1 General troubleshooting 10.1.1 For access Problem Possible causes Remedy No write access to parameters Write protection enabled Check the locking status(→ 38). 10.2 Diagnostics information in DeviceCare and FieldCare 10.2.1 Diagnostic options Any faults detected by the measuring device are displayed on the home page of the operating tool once the connection has been established.
Dosimass Diagnostics and troubleshooting 10.2.2 Calling up remedy information Remedy information is provided for every diagnostic event to ensure that problems can be rectified quickly: • On the home page Remedy information is displayed in a separate field below the diagnostics information.
Page 42
Diagnostics and troubleshooting Dosimass Diagnostic Short text Remedy instructions Status Diagnostic number signal behavior [from the [from the factory] factory] Diagnostic of electronic Software incompatible 1. Check software Alarm 2. Flash or change main electronics module Main electronic failure Change main electronic module F...
The fault cannot be rectified or The following options are available for tackling problems of this nature: some other fault not described • Request the services of an Endress+Hauser service technician. above has occurred. Be ready with the following information when requesting the services...
Diagnostics and troubleshooting Dosimass 10.4.2 Response of outputs to errors The failsafe mode of the pulse, frequency and status output can be set differently via the DeviceCare or FieldCare operating program. Error response mode of the pulse and status output...
Dosimass Diagnostics and troubleshooting Structure of the submenu Diagnostics → Actual diagnostics Previous diagnostics Parameter overview with brief description Parameter Prerequsite Description User interface Factory setting Actual diagnostics 1 diagnostic event has Displays the current diagnostic Symbol for diagnostic –...
Diagnostics and troubleshooting Dosimass • Out of specification (S) • Maintenance required (M) • Information (I) 10.7.3 Overview of information events Unlike a diagnostic event, an information event is displayed in the event logbook only and not in the diagnostic list.
Dosimass Diagnostics and troubleshooting Options Description Restart device The restart resets every parameter whose data are in the volatile memory (RAM) to the factory setting (e.g. measured value data). The device configuration remains unchanged. History reset Every parameter is reset to its factory setting.
"Manufacturer' s information" document. The manufacturer' s information is available: • In the Download Area of the Endress+Hauser Internet site: www.endress.com → Download • Specify the following details: –...
Replacement seals (accessory) (→ 11.2 Measuring and test equipment Endress+Hauser offers a wide variety of measuring and test equipment, such as W@M or device tests. Your Endress+Hauser Sales Center can provide detailed information on the services. For a list of some of the measuring and test equipment, refer to the "Accessories"...
• Spare parts are grouped into logical kits with the associated Installation Instructions. • Repairs are carried out by Endress+Hauser Service or by correspondingly trained customers. • Certified devices can be converted into other certified devices by Endress+Hauser Service or at the factory only. Notes for repair and conversion For repair and modification of a measuring device, observe the following notes: •...
Dosimass Repair Carry out the mounting and connection steps from the chapters "Mounting the measuring device" and "Connecting the measuring device" in the logically reverse sequence. Observe the safety instructions. 12.5.2 Disposing of the measuring device WARNING Danger to personnel and environment from fluids that are hazardous to health.
Various accessories, which can be ordered with the device or subsequently from Endress +Hauser, are available for the device. Detailed information on the order code in question is available from your local Endress+Hauser sales center or on the product page of the Endress+Hauser website: www.endress.com.
Page 53
The application already contains the data of your Endress+Hauser device. Endress +Hauser also takes care of maintaining and updating the data records.
Technical data Dosimass Technical data 14.1 Application The measuring device is suitable for flow measurement of liquids and gases only. Depending on the version ordered, the measuring device can also measure potentially explosive, flammable, poisonous and oxidizing media. To ensure that the device remains in proper operating condition for its service life, use the measuring device only for media against which the process-wetted materials are adequately resistant.
Dosimass Technical data To calculate the measuring range, use the Applicator sizing tool Recommended measuring range "Flow limit" section (→ Operable flow range Over 1000 : 1. Flow rates above the preset full scale value are not overridden by the electronics unit, with the result that the totalizer values are registered correctly.
Technical data Dosimass Number of switching Unlimited cycles Assignable functions • Off • On • Diagnostic behavior – Alarm – Alarm and warning – Warning • Limit value – Mass flow – Volume flow – Density – Temperature • Flow direction monitoring –...
Dosimass Technical data Transmitter and sensor Supply voltage 24 V Nominal voltage (20 to 30 V • The power supply may not exceed a maximum short-circuit current of 50 A. • The measuring device may only be connected to SELV, PELV or CLASS 2 circuits.
Page 58
Technical data Dosimass Zero point stability Zero point stability [mm] [in] [kg/h] [lb/min] ³⁄₈ 0.20 0.007 ½ 0.65 0.024 1.80 0.066 Flow values Flow values as turndown parameter depending on nominal diameter. SI units 1:10 1:20 1:50 1:100 1:500 [mm]...
Dosimass Technical data Mass flow Influence of medium temperature When there is a difference between the temperature for zero point adjustment and the process temperature, the typical measured error of the sensor is ±0.0003 % of the full scale value/°C (±0.00015 % of the full scale value/°F).
Page 60
Technical data Dosimass Seals No internal seals Medium pressure range max. 100 bar (1 450 psi), depending on the process connection (nominal pressure) Density 0 to 5 000 kg/m (0 to 312 lb/cf) Pressure-temperature An overview of the pressure-temperature ratings for the process connections is ratings provided in the "Technical Information"...
Dosimass Technical data 14.10 Mechanical construction Design, dimensions For the dimensions and installation lengths of the device, see the "Technical Information" document, "Mechanical construction" section Weight Compact version Weight in SI units DN [mm] Weight [kg] Weight in US units...
EC Directives. These are listed in the corresponding EC Declaration of Conformity along with the standards applied. Endress+Hauser confirms successful testing of the device by affixing to it the CE mark. C-Tick symbol The measuring system meets the EMC requirements of the "Australian Communications and Media Authority (ACMA)".
NI (non-incendive version) Hygienic compatibility 3A approval Pressure Equipment • With the PED/G1/x (x = category) marking on the sensor nameplate, Endress+Hauser Directive confirms compliance with the "Essential Safety Requirements" specified in Annex I of the Pressure Equipment Directive 97/23/EC.
Page 64
Technical data Dosimass Technical Information Measuring device Documentation code Dosimass TI00065D Supplementary device- Safety Instructions dependent documentation Contents Documentation code ATEX Ex nA XA00079D cCSAus NI FES0232 Endress+Hauser...
Appendix Dosimass Switch output function 1 Switch status 1 Switch output function 2 Switch status 2 Reset overall batching quantity 15.1.2 "Setup" menu Navigation Setup Setup (→ Device tag (→ Device tag (→ ‣ System units (→ Mass flow unit (→...
Page 68
Appendix Dosimass Failure frequency (→ (→ Switch-on value Switch-off value (→ Failure mode (→ Invert output signal (→ ‣ Communication Bus address Baudrate Data transfer mode Parity Byte order Assign diagnostic behavior Failure mode ‣ Low flow cut off (→...
Dosimass Appendix ‣ Sensor adjustment (→ Installation direction (→ ‣ Zero point adjustment (→ (→ Zero point adjustment control Progress (→ ‣ Totalizer 1 to 3 (→ Assign process variable (→ Mass unit (→ Volume unit (→ Totalizer operation mode (→...
Page 70
Appendix Dosimass Timestamp Diagnostics 2 Timestamp Diagnostics 3 Timestamp Diagnostics 4 Timestamp Diagnostics 5 Timestamp ‣ Event logbook Filter options Clear event list ‣ Device information (→ Device tag (→ Serial number (→ Firmware version (→ Device name (→ (→...
Dosimass Appendix Volume flow Density Temperature ‣ Input values Value status input ‣ Output values Pulse output Output frequency Switch status ‣ Totalizer (→ Totalizer value 1 to 3 Totalizer overflow 1 to 3 ‣ Simulation (→ Assign simulation process variable (→...
Page 72
Appendix Dosimass "System" submenu Navigation Expert → System System ‣ Diagnostic handling Alarm delay ‣ Diagnostic behavior Assign behavior of diagnostic no. 140 Assign behavior of diagnostic no. 046 Assign behavior of diagnostic no. 834 Assign behavior of diagnostic no. 835 Assign behavior of diagnostic no.
Page 73
Dosimass Appendix Permanent storage Device tag (→ "Sensor" submenu Navigation Expert → Sensor Sensor ‣ Measured values ‣ Process variables Mass flow Volume flow Density Temperature ‣ Totalizer (→ Totalizer value 1 to 3 Totalizer overflow 1 to 3 ‣...
Page 74
Appendix Dosimass ‣ System units (→ (→ Mass flow unit Mass unit (→ Volume flow unit (→ Volume unit (→ Density unit (→ Temperature unit (→ Date/time format ‣ User-specific units User mass text User mass factor User volume text...
Page 75
Dosimass Appendix Off value low flow cutoff (→ Pressure shock suppression (→ ‣ Partially filled pipe detection (→ Assign process variable Low value partial filled pipe detection (→ High value partial filled pipe detection (→ Response time part. filled pipe detect.
Page 76
Appendix Dosimass Volume flow offset Volume flow factor Density offset Density factor Temperature offset Temperature factor ‣ Calibration Calibration factor Zero point Nominal diameter Calibration date/time Calibration factor ‣ Logbook Calibration counter Calibration date/time Calibration date/time Calibration date/time Calibration date/time...
Page 78
Appendix Dosimass ‣ Testpoints Oscillation frequency Oscillation frequency Frequency fluctuation Frequency fluctuation Oscillation amplitude Oscillation amplitude Oscillation damping Oscillation damping Tube damping fluctuation Tube damping fluctuation Signal asymmetry Exciter current Exciter current RawMassFlow ‣ Supervision Limit value measuring tube damping "Current input"...
Page 79
Dosimass Appendix Active level Response time status input Output ‣ Pulse/frequency/switch output 1 (→ to 2 Operating mode (→ Channel 2 (→ Assign pulse output (→ Value per pulse (→ (→ Pulse width Measuring mode Failure mode (→ Pulse output Assign frequency output (→...
Page 80
Appendix Dosimass Assign limit (→ (→ Switch-on value Switch-off value (→ Assign flow direction check (→ Assign status (→ Failure mode (→ Switch status Invert output signal (→ Communication ‣ Modbus configuration Bus address Baudrate Data transfer mode Parity Byte order...
Page 82
Appendix Dosimass Mapped slot Mapped slot Mapped slot Mapped slot Mapped slot Mapped slot Maximum integer value Mass flow factor Mass flow offset Density factor Density offset Temperature factor Temperature offset Volume flow factor Volume flow offset Mass totalizer factor...
Page 83
Dosimass Appendix ‣ Modbus information Device ID Device revision ‣ Modbus data map Scan list register Scan list register Scan list register Scan list register Scan list register Scan list register Scan list register Scan list register Scan list register...
Page 84
Appendix Dosimass ‣ Totalizer 1 to 3 (→ (→ Assign process variable Mass unit (→ Volume unit (→ Totalizer operation mode (→ Control Totalizer 1 to 3 (→ Preset value 1 to 3 (→ Failure mode (→ ‣ Batching ‣...
Page 85
Dosimass Appendix Switch output function 2 Switch status 2 ‣ Configuration Batch profile ‣ Settings batch profile 1 to 6 Input selector Batch unit Batch quantity Measuring time drip quantity Fixed compensation quantity Drip correction mode Filter depth drip median...
Page 86
Appendix Dosimass Timestamp Actual diagnostics Previous diagnostics (→ Timestamp Previous diagnostics Operating time from restart Operating time ‣ Diagnostic list Diagnostics 1 Diagnostics 1 Timestamp Diagnostics 2 Diagnostics 2 Timestamp Diagnostics 3 Diagnostics 3 Timestamp Diagnostics 4 Diagnostics 4 Timestamp...
Page 87
Dosimass Appendix ‣ Event logbook Filter options Clear event list ‣ Device information (→ Device tag (→ Serial number (→ Firmware version (→ Device name (→ Order code (→ Extended order code 1 (→ Extended order code 2 (→ Extended order code 3 (→...
Page 88
Appendix Dosimass ‣ Min/max values Reset min/max values ‣ Medium temperature Minimum value Maximum value ‣ Simulation (→ Assign simulation process variable (→ Value process variable (→ Simulation device alarm (→ (→ Simulation diagnostic event Endress+Hauser...