Troubleshooting And Maintenance; System Alarms - Ametek 5000 Manual

Moisture analyzer
Table of Contents

Advertisement

TROUBLESHOOTING AND MAINTENANCE

Any troubleshooting or maintenance to this analyzer that requires access to the inte-
!
rior of the components must be performed only by qualified service personnel . There
are no operator-level serviceable parts, expendables, or maintenance needs inside this
WARNING
analyzer.

System Alarms

The system alarm contacts are activated when there is a condition that makes the analyz-
er output questionable. The error messages in Figure 10 will be displayed on the bottom
line of the analyzer display when a system alarm condition exists.
CELL TEMP ALERT
LOW CELL FREQUENCY
HIGH CELL FREQUENCY
NEGATIVE DELTA FREQUENCY
GEN TEMP ALERT
CAL ERROR
DISPLAY OVER-RANGED
MEMORY IS CORRUPTED
Figure 10.
System Alarm error messages.
It is common for some of these system alarms to activate momentarily during normal
operation, especially during equilibration periods between measurement cycles. If the
system alarm persists, there may be a problem with the measuring cell, the moisture
NOTE
generator, or the thermal controller.
Moisture Cell temperature control is suspect
(signal sent from field unit).
Activates when measured cell frequency is < 500 Hz.
Activates when measured cell frequency is > 3500 Hz.
Activates when delta frequency is < –1 Hz.
Moisture Generator temperature control is suspect
(signal sent from field unit).
Activates when the new calculated span value (Moist_Span)
exceeds the limits Max_Cal_Err or Max_Span. Refer to
Chapter 4, "Alarm Button" for more information. Deactivates
after successful calibration.
A displayed value exceeds the allotted display area.
The RAM or RAM battery has failed causing factory default
parameters to be loaded into memory.
Essential Health & Safety Requirements
|
33

Advertisement

Table of Contents
loading

Table of Contents