Activating automatic
synchronization
3. In the Connect Mode drop-down menu, select the option
seca analytics 115 TCP.
The recommended presets are displayed.
4. Adapt the presets if necessary.
5. In the Connect Server IP line, enter the appropriate IP address.
Configuration
PC software
seca analytics 115
as client/server solution
seca analytics 115
PC software
as a standalone solution
6. In the Connect Port line, enter the port of the PC selected under 5. (de-
fault setting: 20010).
You have the following options for continuing:
► Press the

Activating automatic synchronization

► Press the
Confirm settings, exit dialog
If you are using the device in combination with the
software, you should activate automatic synchronization. The following data
are then synchronized as soon as they are changed either on the device or in
the PC software:
• seca patient files
• User accounts and passwords
• Automatic export settings
NOTICE!
Malfunction
In order to be able to use automatic synchronization, a connection to
the PC software must be set up ➔
seca analytics 115 PC
► Ensure that the connection to the PC software is set up before you
activate automatic synchronization.
IP address
The IP address of the PC on which
the
seca analytics 115
was installed with the Server or
options
Complete
IP address of the PC workstation
button:
button:
seca analytics 115
Setting up a connection to the
software.
Setting up peripherals • 175
English
PC software
PC