Vital Signs Measurement; General; Blood Pressure Measurement - Seca mVSA 535 Instructions For Use Manual

Hide thumbs Also See for mVSA 535:
Table of Contents

Advertisement

11.4 Vital signs measurement

General

Fault
Vital signs tab is not
displayed
is being
seca mVSA 535
operated on a wheeled
stand: Measuring results
for vital signs are
implausible

Blood pressure measurement

Fault
Implausible measuring
results
Insufficient cuff pressure
No option to select
Multiple measurement in
the NIBP settings dialog
window
Blood pressure values do
not appear in the EMR
system or in the
PC
seca analytics 115
software
112 •
/
seca mVSA 535
seca mBCA 525
General
Blood pressure measurement
COVIDIEN
COVIDIEN
Manual temperature input
SpO
measurement
2
Cause
USB connecting cable of the
SmartBucket not connected to the
monitor
SmartBucket not activated
• Original seca USB connecting cable
not being used
• USB extension cable also in use
Cause
Excessive patient movement
Incorrect blood pressure cuff used
Blood pressure cuff not applied
correctly
Blood pressure cuff applied to an
extremity to which there is intravenous
access
Incorrect blood pressure cuff used
Blood pressure cuff or compressed air
tube leaking
Pump in device faulty
Administrator has configured a
measuring profile. The name of the
measuring profile is displayed.
Multiple measurement was performed
TM
TM
FILAC
3000 temperature measurement
TM
®
GENIUS
3 temperature measurement
• Connect the USB connecting cable of the
SmartBucket to the monitor
• Restart the device
Administrator:
Activate SmartBucket: ➔
system components
menu
• Use original seca USB connecting cable
(in scope of delivery for the wheeled stand)
• Do not use a USB extension cable
• Ask the patient to move as little as possi-
ble
• Use correct size of blood pressure cuff
• Only use seca blood pressure cuffs
Correctly apply blood pressure cuff, see
instructions for use for blood pressure cuff
Apply blood pressure cuff to a different
extremity
• Use correct size of blood pressure cuff
• Only use seca blood pressure cuffs
Dispose of blood pressure cuff, use
replacement
Do not continue using device and have it
repaired by seca Service
Not a malfunction
• Press the name of the measuring profile
• ➔
Starting a multiple
For current measurement:
• ➔
Modifying presets
single measurement
Administrator:
• Specify Individual measurement as preset
Presets for blood pressure
Remedy
Managing
in the administrator
Remedy
measurement.
and ➔
Starting a