seca SpO
measurement
2
Measurement
Measuring method
LED wavelength:
Red
Infrared
Maximum light output
This information may be of particular interest to clinicians
Measuring range:
• SpO
2
• Pulse rate (default)
• Pulse rate (enhanced)
Measuring accuracy:
• SpO
(no patient movement)
2
• SpO
(patient movement)
2
• Pulse rate (no patient movement)
• Pulse rate (patient movement)
Resolution:
• Oxygen saturation (SpO
• Pulse rate
IEC 60601-1: Medical electrical device, type BF
a.Validated by clinical tests in which the sensor values measured were compared to those of the arterial CO oximetry of healthy
adults over the specified range of functional oxygen saturation.
b.As the measurements of pulse oximeters are statistically distributed, only approximately two-thirds of these measurements
will probably fall in the range from ± A
c.Tested with a Fluke Index II Oximeter Tester (all movement patterns).
128 •
/
seca mVSA 535
seca mBCA 525
No Implied License Statement
Possession or purchase of this device does not convey any express or
implied license to use the device with unauthorized sensors or cables which
would, alone or in combination with this device, fall within the scope of one or
more of the patents relating to this device.
seca SpO
)
2
(accuracy root mean square) of the value measured by CO oximeters.
RMS
a b
module
2
Functional oxygen saturation
Spectrophotometry (red/infrared)
660 nm
900 nm
0 % – 100 %
30 min
– 240 min
-1
20 min
– 300 min
-1
60 % – 70 % ± 3 A
60 % – 100 % ± 2 A
<60 % not specified
70 % – 100 % ± 3 A
<70 % not specified
≤ 2 min
5 mW
-1
-1
rms
rms
c
rms
-1
-
1 %
1 min
-1