Fault
BIA tab not active
Patient and user data can
not be restored manually
Patient and user data
cannot initially be
encrypted after software
update
11.2 Measuring mat
Fault
Measuring mat cannot be
switched on
Charge status LED is red
WiFi LED is red
One or more Electrode
contact LEDs are red
110 •
/
seca mVSA 535
seca mBCA 525
Cause
No seca patient file prepared
Measuring mat not activated
No measuring mat present
USB stick contains several backup
copies
Backup copy on the USB stick has
been renamed
There is not enough memory on the
monitor for initial encryption
Cause
Rechargeable battery discharged
Measuring mat faulty
Rechargeable battery discharged
Rechargeable battery faulty
Inductive charging interface of monitor
covered, e.g. by other measuring
accessories
Inductive charging interface faulty
No WiFi connection to monitor
WiFi module of measuring mat faulty
Electrode cables not connected to
electrodes
Electrodes defective
Electrode cables or measuring mat
faulty
Remedy
Prepare seca patient file ➔
bioimpedance measurement
Administrator:
Activate measuring mat: ➔
system components
in the administrator
menu
Administrator, if desired:
• Retrofit measuring mat
• Activate measuring mat: ➔
system components
in the administrator
menu
Ensure that there is only one folder on the
USB stick with the designation
<ddMMyyyy_hhmm>_seca_monitor_b
ackup (if necessary, rename other folders)
Ensure that backup copies are saved in the
folder with the designation
<ddMMyyyy_hhmm>_seca_monitor_b
ackup (if necessary, rename folders)
Initial manual encryption by administrator:
➔
Exporting patient and user data
manually
➔
Resetting the device
➔
Restoring patient and user data
manually
• If necessary, make previous device
settings (for example, WiFi connection)
manually
Remedy
Suspend measuring mat in storage
compartment of monitor and charge
rechargeable battery
Replace measuring mat
Suspend measuring mat in storage
compartment of monitor and charge
rechargeable battery
Replace measuring mat
First suspend the measuring mat in the
magnetic catch of the monitor, then stow
other measuring accessories
Inform seca Service
Administrator:
Check settings for WiFi connection and
correct if necessary
Replace measuring mat
Ensure that all electrode cables are
engaged in the push-buttons of the
electrodes
Replace electrodes
Replace measuring mat