4. Press the
button.
The Configure multiple measurement dialog window appears:
5. Specify the number of part-measurements:
• Maximum six part-measurements
• Deactivated part measurements at the start and during the measuring
sequence are performed but not used for forming the mean value
• Deactivated part-measurements at the end of the measuring sequence
are not completed: The running time of the measurement is reduced.
Example
6. Specify a start delay (waiting time before part-measurement 1):
• Format: mm:ss
• Min: 00:00, Max:10:00
7. Specify the interval (waiting time between part-measurements)
• Format: mm:ss
• Min: 01:35, max: 14:15 minutes
Measuring sequence
Factory setting: 6 part-
measurements
Part-measurement 1 is discarded
Example: 6 part-measurements
Part-measurements 1, 4 and 5 are
discarded
Example: 4 part-measurements
• Part-measurement 1 is discarded
• Part-measurements 5 and 6 are not
performed
Making settings for measuring mode • 151