11.7 EMR system connection via seca connect 103 software

Fault
Data transmission cannot
be set up between device
and
seca connect 103
Unable to find user ID
Unable to find patient ID
Unable to scan IDs
There is no request for a
password following input
of the EMR user ID. The
Vital signs tab is displayed
immediately
Unable to find EMR system
patient data
Login using EMR user ID
not possible
Blood pressure values do
not appear in the EMR
system
Not possible to submit
measuring results to the
EMR system
118 •
/
seca mVSA 535
seca mBCA 525
For more information on using the
seca 103/452 system instructions for use.
Cause
Software versions not compatible
BIA measuring mat activated.
EMR system connections are possible
only with the BIA measuring mat
deactivated
No connection to the server
User ID not recognized by the server
No connection to the server
Patient ID not recognized by the server
Scanner not connected correctly
Barcode scanner incompatible
Scanner defective
Not a malfunction:
The device connection to the EMR
system has been configured so that no
password needs to be entered.
No patient file available in the EMR
system
Error scanning patient ID
No network connection to the EMR
system
EMR system connection interrupted,
device set up so that no measurements
are possible
Multiple measurement was performed
Network cable not plugged in correctly
Network cable defective
Device outside WiFi range
No network connection to the EMR
system
Network connection interrupted
software, see the
seca connect 103
Remedy
Administrator:
Use a compatible version of the
software ➔
seca connect 103
Deactivate measuring mat, more
information is available here: ➔
the measuring mat (seca mVSA 535 only)
Administrator:
Check server connection
Ensure that user ID is valid
Administrator:
Check server connection
Ensure that patient ID is valid
Ensure that the USB stick is correctly
located in the USB interface
Use compatible barcode scanner
Replace scanner
Administrator:
Modify configuration if necessary
Making workflow settings
Create patient file directly in the EMR
system
Scan patient ID again
Administrator:
Check the network connection and set up
again if necessary
Administrator:
If possible, allow measurements with the
EMR system connection interrupted
Making workflow settings
For current measurement:
• ➔
and ➔
Modifying presets
single measurement
Administrator:
• Specify Individual measurement as preset
Presets for blood pressure
Check whether the network cable is
connected correctly and that the
connectors are firm in their sockets
Replace network cable
Bring device into WiFi range
Administrator:
Check the network connection and set up
again if necessary
If network connection is active ➔
Unsubmitted measurements list
Compatibility
Retrofitting
Starting a
Using the