Handling A Wheeled Stand - Seca mVSA 535 Instructions For Use Manual

Hide thumbs Also See for mVSA 535:
Table of Contents

Advertisement

Handling a wheeled stand

12 •
/
seca mVSA 535
seca mBCA 525
CAUTION!
Patient hazard, damage to device
► Additional devices which are connected to electrical medical devic-
es must provide evidence of compliance with the relevant IEC or ISO
standards (e.g. IEC 60950 for data-processing devices). Further-
more, all configurations must comply with the requirements of stan-
dards for medical systems (see IEC 60601-1-1 or Section 16 of
edition 3.1 of IEC 60601-1, respectively). Anyone connecting addi-
tional devices to electrical medical devices is considered a system
configurer and is therefore responsible for ensuring that the system
complies with the requirements of standards for systems. Your at-
tention is drawn to the fact that local laws take precedence over the
above-mentioned requirements of standards. In the event of any
queries, please contact your local specialist dealer or Technical
Service.
► Please have servicing and measuring technology checks performed
every two years.
► Technical modifications may not be made to the device. The device
does not contain any parts for servicing by the user. Only have ser-
vicing and repairs performed by an authorized seca Service partner.
You can find service partners in your area at www.seca.com or by
sending an e-mail to service@seca.com.
► Only use original seca accessories and spare parts, otherwise seca
will not grant any warranty.
CAUTION!
Patient hazard, malfunction
► Keep other electrical medical devices, e.g. high-frequency surgical
devices, a minimum distance of approx. 1 meter away to prevent in-
correct measurements or wireless transmission interference.
► Keep HF devices such as cell phones a minimum distance of ap-
prox. 1 meter away to prevent incorrect measurements or wireless
transmission interference.
► The actual transmission output of HF equipment may require mini-
mum distances of more than 1 meter. Details can be found at
www.seca.com.
WARNING!
Injury from falling, damage to device
► When transporting the device on a wheeled stand, make sure that
all cables and tubes are stowed properly directly on the machine or
in the basket of the stand.
CAUTION!
Damage to device
► Do not pull on cables and tubes to move the equipment or wheeled
stand.
► Do not move the wheeled stand when the power cord of the device
is plugged into an electrical outlet.