Disinfecting; Sterilizing - Seca mVSA 535 Instructions For Use Manual

Hide thumbs Also See for mVSA 535:
Table of Contents

Advertisement

8.2 Disinfecting

Component (depending on equipment)
: Monitor with storage
seca mBCA 525
compartment
: Monitor with
seca mVSA 535
SmartBucket
Measuring mat and electrode cables
Blood pressure cuff and compressed air
tube
Temperature probes (red/blue) with cable
Probe holder (red/blue)
In-ear thermometer, cable, measuring
head and lens
Magazine holder for probe covers (in-ear
thermometer)
SpO
sensor with cable
2
Patient cable for SpO

8.3 Sterilizing

104 •
/
seca mVSA 535
seca mBCA 525
1. Follow the instructions for use for the disinfectant.
2. Disinfect the device and its accessories as described in the table:
Interval
As required
Before
and after a
measurem
ent
As required
As required
As required
As required
As required
Follow the manufacturer's instructions for use
sensor
Follow the manufacturer's instructions for use
2
The device and accessories must not be sterilized.
Disinfecting
1. Remove all the measuring accessories (measuring
devices and consumables) from the device (depending
on version)
• ➔
Removing the probe holder
• ➔
Removing the magazine holder
2. Moisten a soft cloth with disinfectant (active ingredient:
quaternary ammonium compounds)
3. Wipe over all surfaces
4. Allow to air-dry for approx. 30 minutes
1. Moisten a soft cloth with disinfectant (active ingredient:
quaternary ammonium compounds)
2. Wipe over measuring mat and electrode cables
3. Allow to air-dry for approx. 30 minutes
1. Moisten a soft cloth with disinfectant (active ingredient:
70 % isopropanol)
2. Wipe over cuff and compressed air tube
3. Rinse thoroughly with water
4. Allow to dry at room temperature
1. Discard probe cover and dispose of it
2. Moisten a soft cloth with disinfectant (active ingredient:
70 % isopropanol)
3. Wipe over temperature probe
4. Shake out temperature probe so that no liquid remains
in it
5. Allow to air-dry for approx. 30 minutes
1. ➔
Removing the probe holder
2. Moisten a cotton bud with disinfectant (active
ingredient: 70 % isopropanol)
3. Wipe over the inner surfaces of the probe holder
4. Allow to air-dry for approx. 30 minutes
1. Discard probe cover and dispose of it
2. Wipe over in-ear thermometer several times with a
cloth moistened with alcohol (70 % isopropanol) so that
all surfaces are visibly moist for at least one minute
3. Make sure that the lens at the tip of the measuring
head does not have any fingerprints or marks on it
4. Allow the in-ear thermometer to dry completely in air
1. ➔
Removing the magazine holder
2. Moisten a cotton bud with disinfectant (active
ingredient: quaternary ammonium compounds)
3. Wipe over surfaces
4. Allow to air-dry for approx. 30 minutes