8. HYGIENIC TREATMENT OF THE SECA 452 INTERFACE MODULE

8.1 Cleaning

8.2 Disinfecting

8.3 Sterilizing

72 •
/
seca connect 103
seca 452
Cleaning
Disinfecting
Sterilizing
WARNING!
Electric shock
The device is not de-energized when the on/off key is pressed and the
display goes out. Use of fluids on the device may cause electric
shock.
► Ensure that the device is switched off before performing any hygiene
treatment.
► Disconnect the power supply connector before performing any
hygiene treatment.
► Before each hygiene treatment, take the rechargeable battery out of
the device (if present and removable).
► Ensure that no fluids penetrate the device.
CAUTION!
Damage to device
Inappropriate detergents and disinfectants may damage the sensitive
surfaces of the device.
► Use only disinfectants free of chlorine and alcohol which are explic-
itly suitable for acrylic sheet and other sensitive surfaces (active in-
gredient: quaternary ammonium compounds, for example).
► Do not use caustic or abrasive detergents.
► Do not use organic solvents (e.g. white spirit or petroleum spirit).
► Use a soft cloth dampened with mild soapsuds to clean the surfaces of
the device.
1. Check that your disinfectant is suitable for sensitive surfaces and acrylic
sheet.
2. Follow the instructions for use for the disinfectant.
3. Disinfect the device by moistening a soft cloth in disinfectant and wiping
down the device.
This device may not be sterilised.