Appendices
Error while charging the product
LED on
Status LED
power sup
on battery
ply
charger
Status LED Charging status indicat
The
LED
ring is lit in
weak purple
The
LED
ring is lit up
yellow
The
LED
ring is lit up
green
The
LED
ring is flash
ing red
92 | Ottobock
Error
Country-specific plug adapter not fully
engaged on power supply
Non-functional socket
Defective power supply
No connection between battery charger
and power supply
Defective battery charger
or (5 LEDs)
No LED is lit
LED 2 and 4 are lit up
LED 1, 3 and 5 are lit up
Resolution
Check whether the country-specific plug
adapter is fully engaged on the power
supply.
Check socket with another electric appli
ance.
The battery charger and power supply
must be inspected by an authorised
Ottobock Service Centre.
Check whether the charging cable plug is
fully engaged on the battery charger.
The battery charger and power supply
must be inspected by an authorised
Ottobock Service Centre.
Error
Distance between battery
charger and receiver of the
charging unit on the pros
thesis too great. If the dis
tance is more than 2 mm,
the prosthesis cannot be
charged.
Charger excess temperat
ure
Excessively high or low
temperature of the pros
thesis
The prosthesis is not being
charged
Distance between battery
charger and receiver of the
charging unit too great.
The
battery
charger
functional but not connec
ted to the receiver, or the
distance between the bat
tery charger and receiver
of the charging unit is too
great.
The prosthesis is not being
charged
Defective battery charger.
Resolution
Reduce distance between
battery charger and receiver
of the charging unit.
Check whether the spe
cified ambient conditions for
charging the battery are met
(see Page 94).
Connection
may
improved by reducing the
distance between the char
ger and receiver of the char
ging unit.
is
Connect the battery charger
or
reduce
the
distance
between the charger and
the receiver of the charging
unit on the prosthesis.
Reset the error by discon
necting and connecting the
power supply.
If the error persists, the bat
tery charger and power sup
ply must be inspected by an
authorised Ottobock Ser
vice Centre.
Genium 3B1/3B1=ST
be