Table 5-8 Command Sequence For Memory Patterns Test Example Using Individual Memory Pattern Recall - Keithley 2790 User Manual

Sourcemeter switch system
Hide thumbs Also See for 2790:
Table of Contents

Advertisement

Model 2790 SourceMeter
Table 5-8
Command sequence for memory patterns test example using individual memory
pattern recall
*RST
ROUT:MEM:CLE:ALL
Set up memory location 1:
ROUT:MEM:CHAN 1,(@101,114,118)
ROUT:MEM:DEL 1,0
ROUT:MEM:READ:STAT 1, ON
SENS:FUNC 'RES',(@M1)
SENS:RES:RANG 100,(@M1)
SENS:RES:NPLC 1,(@M1)
Set up memory location 2:
ROUT:MEM:CHAN 2,(@104,114,118)
ROUT:MEM:DEL 2,0
ROUT:MEM:READ:STAT 2, ON
SENS:FUNC 'RES',(@M2)
SENS:RES:RANG 100,(@M2)
SENS:RES:NPLC 1,(@M2)
Set up memory location 3:
ROUT:MEM:CHAN 3,(@108,115,118)
ROUT:MEM:DEL 3,0
ROUT:MEM:READ:STAT 3,ON
SENS:FUNC 'RES',(@M3)
SENS:RES:RANG 100,(@M3)
SENS:RES:NPLC 1,(@M3)
FORM:ELEM READ
Recall memory locations, trigger/request readings:
ROUT:MEM:REC 1
READ?
ROUT:MEM:REC 2
READ?
ROUT:MEM:REC 3
READ?
®
Switch System User's Manual
Command Sequence
Description
Restore GPIB defaults.
Clear all memory pattern locations.
Set channels 1, 14, and 18 for memory location 1.
Set delay to 0.
Enable reading.
Select 2-wire ohms.
Select 100Ω range.
Select 1 PLC.
Set channels 4, 14, and 18 for memory location 2.
Set delay to 0.
Enable reading.
Select 2-wire ohms.
Select 100Ω range.
Select 1 PLC.
Set channels 8, 15, and 18 for memory location 3.
Set delay to 0.
Enable reading.
Select 2-wire ohms.
Select 100Ω range.
Select 1 PLC.
Send readings only from buffer.
Recall memory pattern 1.
Trigger and request reading.
Recall memory pattern 2.
Trigger and request reading.
Recall memory pattern 3.
Trigger and request reading.
5-33

Hide quick links:

Advertisement

Table of Contents
loading