Position the patient within 18 inches (45 cm) of the Remote Control and orient
the large surface of the Remote Control parallel to large surface of the Stimulator.
Proper orientation will ensure a reliable communication link between the Clinician
Programmer and the Stimulator.
Note: A message will be displayed on the Clinician Programmer if the
communication range is not satisfactory.
7
Turn on the Clinician Programmer and log in.
8
If only Bionic Navigator is installed on the Clinician Programmer (CP), click on
the Bionic Navigator icon
If more than one DBS software is installed on the CP, Click the Vercise Launcher
icon
and select the Vercise DBS System icon
Note: Multiple programming softwares should not be run simultaneously on the
same CP.
9
Press and hold the
Note: The Remote Control should be linked to the Stimulator (see the "Linking a
Remote Control to a Stimulator" section of this manual).
10
Enter the Clinician's screen on the Remote Control by pressing and holding both
P
the and
buttons simultaneously.
P
11
Press the
button to select "CP Mode."
The Remote Control will inform you that it is ready to communicate with the
Clinician Programmer.
Once the Remote Control is within range of the Stimulator, the Clinician Programmer
will detect the Stimulator and begin communication.
to launch Bionic Navigator.
P
button to unlock the Remote Control.
Setting Up
to start Bionic Navigator.
Vercise™ DBS Programming Manual