Setting Up

Connecting the Remote Control to the Clinician
Programmer
The Clinician Programmer communicates with the Stimulator via the Remote Control.
The Clinician Programmer is connected to the Remote Control using an infrared (IR)
link. The Remote Control uses a radio frequency (RF) link to communicate with the
Stimulator. Programs can be loaded to the Clinician Programmer for previewing,
activating, and changing stimulation parameters. Stimulation programs can also be saved
to the Remote Control.
CAUTION: Use only Vercise™ DBS System components with the Clinician
Programmer. Failure to do so can result in inappropriate stimulation.
CAUTION: The CP is not equipment for the patient environment as defined by
IEC60601‑1. The CP and the person using the CP should not be in
contact with the patient while programming.
To connect the remote Control to the Clinician Programmer:
1
Power on the CP.
2
Log in as ClinicUser. If this is the first login, use "bsn" as a temporary password.
You will be prompted to setup a new ClinicUser password when logging into the
CP for the first time. The new ClinicUser password must be 10 or more characters
in length.
Note: In regions where software is installed by a third party, follow region‑specific
guidance for the initial ClinicUser password.
3
Plug the USB to Serial Cable into the USB port on the Clinician Programmer.
4
Plug the serial end of the USB to Serial Cable into the matching end of the IR
Interface.
5
Insert the IR Interface and Remote Control into the plastic holder to keep the IR
windows aligned, as shown below.
1
Vercise™ DBS Programming Manual
91053785-17
2 of 1705
2
3