Configuring The Vercise Dbs System; Verifying The Vercise Dbs System Configuration - Boston Scientific Vercise DBS Programming Manual

Hide thumbs Also See for Vercise DBS:
Table of Contents

Advertisement

Vercise™ DBS Programming Manual
Configuring the Vercise DBS
System

Verifying the Vercise DBS System Configuration

The Configuration (or Config) screen displays information about the system components.
The Stimulator and Remote Control product and serial numbers are automatically
detected and recorded. The Lead information is also displayed.
Note: The Configuration screen may only be accessed once a patient is selected.
To verify the Vercise DBS System configuration:
1
Click
If Lead information has not been entered, the information will be requested before
the Configuration screen is displayed.
2
Verify that the Lead information is correct.
3
Verify that the Stimulator model and serial number are correct.
4
Verify that the Remote Control model and serial number are correct.
5
Verify that the electrode labels correctly correspond to the left and right Leads
(in the image of the Leads to the right of the screen).
By default, Bionic Navigator assigns contacts 1 through 8 to the left Lead
and contacts 9 through 16 to the right Lead. If the contacts do not correspond
correctly, see the "Swapping the Leads" section of this manual.
Vercise™ DBS Programming Manual
91053785-17
16 of 1705
in the upper right‑hand corner of any screen.

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents