Programming Multiple Areas

Programming Multiple Areas with the Same Rate
The Vercise™ DBS System allows two Areas to be programmed with the same rate (or
frequency). If more than two Areas are turned on, then all Areas must have the same rate.
Note: The rates for the first and second Areas must be equal to one another before a third
Area may be turned on.
Programming Two Areas with Different Rates
The Vercise DBS System allows two Areas to be programmed with different rates (or
frequencies). This feature allows you to simultaneously stimulate two areas of the brain
with different rates.
To program two Areas with different rates:
1
Turn on one Area and turn off all the other Areas.
2
Select the desired rate for the first Area.
3
Turn on a second Area.
4
Select the desired rate for the second Area.
Note: You will have a limited number of rates available for programming when
selecting the rate for the second Area.
Note: Modifying the rate of either Area will alter the available rates for the other
Area.
Note: When establishing two areas with different rates, the first area must be on
while the rate for second area is programmed.