Adjusting Patient Options

The Patient Options window allows you to view and edit additional program settings,
including stimulation cycling, ramp up time, and Remote Control amplitude lockout.
Program settings are only applied to the currently active program (which appears on
the left side of the Remote screen). After making changes to Patient Options, save the
currently active program to a slot to preserve the changes.
To access Patient Options:
1
Click
from any screen or
2
Click
on the left side of the Remote screen.
Changing the Stimulation Cycle
The Cycle option determines the on and off duration of stimulation delivery.
For continuous stimulation, the Cycle option should be unchecked.
The option is unchecked by default, resulting in continuous stimulation.
For intermittent stimulation:
1
Check the box next to the Cycle option.
2
Set the stimulation on time using the
The range for stimulation on time is 1 second to 90 minutes.
3
Set the stimulation off time using the
The range for stimulation off time is 1 second to 90 minutes.
Programming the Patient
from the Programming screen.
buttons next to the On field.
buttons next to the Off field.
Vercise™ DBS Programming Manual
91053785-17 37 of 1705