Testing Impedances

The Impedance button is used to verify electrical integrity. Lead impedance is measured
and displayed for each of the Stimulator's 16 contacts. Impedances over 4500 Ω are
considered to be resultant from open or unconnected wires and are displayed with an
orange bar on the Impedance screen (or an "X" on the Program screen).
To check the Lead impedances:
1
Click
while in the Configuration screen or the Program screen.
The impedances will be measured and displayed. To re‑measure the impedances,
click
Note: All measured impedances are saved into the Clinician Programmer
database.
2
To print the impedance report, click
3
To exit, click
.
.
Configuring the Vercise DBS System
.
Vercise™ DBS Programming Manual
91053785-17 19 of 1705