Figure 5-18 Summary Of Standard Event Status Register; Table 5-10 Standard Event Status Register - Xantrex Watt Series Programmable DC Power Supply XG 850 Operating Manual

Xg 850 watt series programmable dc power supply
Table of Contents

Advertisement

Remote Operation
Standard Event
Status Register
(SESR)
Operation Complete
0
Not Used
1
2
Query Error
Device Dependent Error
3
...
Execution Error
4
Command Error
5
Not Used
6
Power On
7

Figure 5-18 Summary of Standard Event Status Register

Table 5-10 Standard Event Status Register

Bit
Bit Weight Bit Name
0
1
1
2
2
4
3
8
4
16
5-32
Figure 5-18 summarizes the Standard Event Status Register.
SESR
Summary
Bit
QUEStionable Status Summary Bit
+
OPERation Status Summary Bit
Operation Complete
(OPC)
Request Control (RQC)
Query Error (QYE)
Device Dependent Error
(DDE)
Execution Error (EXE)
Status Byte
Register
Not Used
Not Used
Error/Event Queue Status Flag
Not Used
Description
Set if *OPC command has been received and all
pending operations have been completed.
Not implemented. Always set to 0.
Not implemented.
Set if there is a device-specific error.
See "Device-Specific Error List" on page B–5
for a list possible error codes and descriptions.
Set if a program data element, following a header,
was evaluated by the power supply as outside of its
legal input range, or is otherwise inconsistent with
the power supply's capabilities. Suggests that a
valid program message could not be properly
executed due to some power supply condition.
See "Execution Error List" on page B–4 for a list of
possible error codes and descriptions.
MSS
Summary
Bit
0
1
2
3
+
4
5
6
7