Remote Operation

Table 5-11 OPERation Status Register

Bit
Bit Weight Bit Name
0
1
1
2
2
4
3
8
4
16
5
32
6
64
7
128
8
256
9
512
10
1024
11
2048
12
4096
13
8192
14
16384
5-36
CALibrating
SETTling
RANGing
SWEeping
MEASuring
Waiting for Arm
Waiting for Trigger
CORRecting
Not Used
SHUTdown Summary
Local Lockout
Current Share Summary Reflects the summary of the CSHare Sub-Register.
Constant Voltage Mode
Constant Current Mode
Program Running
The operation shutdown status register describes the cause of the power
supply shutting down the output. More than one bit may be active and
multiple actions will be required to restart the unit. The protection
shutdown sub-register indicates which protection mechanisms have
caused the power supply to shutdown.
Description
Indicates that the supply is in CALibration Mode.
Not implemented
Not implemented
Not implemented
Not implemented
Not implemented
Not implemented
Not implemented
Not Used
Reflects the summary of the SHUTdown Sub-
Register.
Front panel lockout
Reflects the output mode of the supply.
Reflects the output mode of the supply.
Indicates if an Autosequence program is currently
running.
M370078-01