Step 1: Gain Calibration; Step 2: Offset Calibration; Figure 6-2 Calibration: Step 1 Gain Calibration; Figure 6-3 Calibration: Step 2 Offset Calibration - Xantrex Watt Series Programmable DC Power Supply XG 850 Operating Manual

Xg 850 watt series programmable dc power supply
Table of Contents

Advertisement

Calibration and Troubleshooting

Step 1: Gain Calibration

Step 2: Offset Calibration

6-4

Figure 6-2 Calibration: Step 1 Gain Calibration

Adjust the gain so that the real line and ideal line intersect at a
programmed value of 90%.

Figure 6-3 Calibration: Step 2 Offset Calibration

Adjust the offset so that the real and ideal lines intersect at a programmed
value of 10%.
M370078-01