Table 3-9 Power Supply Default Settings - Xantrex Watt Series Programmable DC Power Supply XG 850 Operating Manual

Xg 850 watt series programmable dc power supply
Table of Contents

Advertisement

Local Operation
Hard Reset
3-40

Table 3-9 Power Supply Default Settings

Parameter
UVP
Foldback trigger
Foldback delay
Current Share Mode
Alarm Output Latches
Alarms Mask
Interlock
Voltage Analog Programming
Voltage APG Scale
Current Analog Programming
Current APG Scale
The hard reset performs the same functions as a soft reset, but will also
clear all calibration constants. Do not perform a hard reset unless you are
planning to re-calibrate all settings of your power supply.
To reset the power supply:
1. Turn the front panel power switch to the Off position.
2. Allow the unit to power down.
3. When the fans are off and the front panel displays are not illuminated,
turn the power switch to the On position.
4. Press and hold the OUTPUT ENABLE Main and OUTPUT
ENABLE Aux buttons.
The front panel should display P5U S rS.
5. When the display reads P5u Clr, release the OUTPUT ENABLE
Main and OUTPUT ENABLE Aux buttons. After 1 second the model
number will appear on the display for 1 second.
The power supply has been reset.
Setting
0 V
None
0.5 s
MASTer
263 (0 × 107, all latches enabled)
2047 (0 × 7FF, all enabled)
Disabled
Off
10 V
Off
10 V
M370078-01