Voltage-Controlled Current Isol Setup - Xantrex Watt Series Programmable DC Power Supply XG 850 Operating Manual

Xg 850 watt series programmable dc power supply
Table of Contents

Advertisement

Analog Programming (APG) and Isolated Analog Programming (ISOL)

Voltage-Controlled Current ISOL Setup

Activating ISOL Programming Current Mode
Deactivating ISOL Current Mode
4-26
1. Turn the 9-position mode control to the CAP position or press the
rotary adjust/Enter control if the control knob is already at the CAP
position.
CAPr is displayed on the output voltage display.
2. Turn the rotary Adjust/Enter control to select U IS (voltage source
from isolated connector).
3. Press the rotary adjust/Enter control to commit the selected mode.
LE C is displayed on the output voltage display.
4. Select the input source level using the rotary adjust/Enter control.
5. Once the desired level has been selected, press the rotary adjust/
Enter control to commit the change.
The display will return to the normal display.
To deactivate the ISOL current mode:
1. Turn the 9-position mode control to the VAP position or press the
rotary Adjust/Enter control if the control knob is already at the VAP
position.
2. CAPr is displayed on the output voltage display.
3. Turn the rotary Adjust/Enter control to select nonE.
4. Press the rotary Adjust/Enter control to commit the change.
The display will return to the normal display.
M370078-01