CURRent
This shows whether the present current level is over or under the specified trip limit.
Sub-Register
Table 4.9 CURRent Sub-Register
Bit
Standard
The standard event status register sets bits for specific events during power supply
operation. All bits in the standard event status registers are set through the error event
Event Status
queue. The register is defined by 488.2 and is controlled using 488.2 common
Register
commands, *ESE, *ESE?, and *ESR?.
Figure 4.3
Device Dependent Error
Figure 4.3 IEEE 488.2 Status Register and Status Byte
Release 2.1
Bit Weight
0
1
1
2
summarizes the standard Event Status Register
Operation Complete
Not Used
Query Error
Execution Error
Command Error
User Request
Power On
Reserved
Reserved
Reserved
Reserved
Reserved
Reserved
Reserved
Reserved
Standard Event Status Register
Bit Name
Description
Over CURrent
Set if the supply's output current is
greater than the user-specified
over-current trip level (variable trip limit)
and the supply is in Operation state.
Under CURrent
Set if the supply's output current is less
than the user-specified under current
trip level (variable trip limit) and the
supply is in Operation state.
0
1
2
3
4
5
6
7
+
8
9
10
11
12
13
14
15
Status Registers
Not Used
Not Used
Not Used
Summary of QUESTionable Status
RQS/MSS
Summary of OPERation Status
Status Byte
Overview
0
1
2
3
MAV
4
5
6
7
77