Isolated Current Monitoring Calibration

Important:
calibration procedure again but use the intended operating voltage level in steps 3
and 8.
To calibrate the isolated current monitoring:
1. Short the main output with a shunt. Connect a multimeter to measure
2. Connect the voltmeter to the Isolated monitoring output lines (J3.10
3. Set the current APG level to 4 V. The SCPI command is:
4. Set the input source to local for voltage analog programming. the
5. Set the main output voltage to 100% by sending the command:
6. Turn the main output to On by pressing the OUTPUT ENABLE Main
7. Enter the interactive calibration mode by sending the SCPI command:
8. Press the terminal keys – or + to reduce or increase the monitoring
9. Press any key other than the + or – keys to exit from interactive
M370078-01
For maximum accuracy at a specific APG level. Follow the
the voltage across the shunt.
and J3.6).
SYST:COMM:APR:LEV:CURR 4
SCPI command is:
SYST:REM:SOUR:CURR LOC
SOUR:VOLT
<ModelVolt>
or use the rotary Adjust/Enter control.
button on the front panel or send the command:
:OUTP ON
The OUTPUT ENABLE Main button will illuminate.
CAL:MON:CURR:ISOL
value. Adjust the monitoring value displayed on the voltmeter that
was set up in Step 1 to read as close as possible to 4.000 V.
calibration mode.
Calibration and Troubleshooting
6
6-17