Analog Programming (APG) and Isolated Analog Programming (ISOL)

Voltage-Controlled Voltage ISOL Setup

Activating ISOL Programming Voltage Mode
Deactivating ISOL Voltage Mode
4-24
To activate ISOL programming voltage mode with an external
voltage source:
1. Turn the 9-position mode control to the VAP position or press the
rotary Adjust/Enter control if the mode control is already at the
VAP position.
WAPr is displayed on the output voltage display.
2. Turn rotary Adjust/Enter control to select U IS (voltage source
from isolated connector).
3. Press the rotary Adjust/Enter control to commit the selected mode.
LE U is displayed on the output voltage display.
4. Select the input source level using the rotary Adjust/Enter control.
5. Once the desired level has been selected, press the rotary Adjust/
Enter control to commit the change.
The display will return to the normal display.
To deactivate the ISOL voltage mode:
1. Turn the 9-position mode control to the VAP position or press the
rotary Adjust/Enter control if the mode control is already at the
VAP position.
2. UAPr is displayed on the output voltage display.
3. Turn the rotary Adjust/Enter control to select nonE.
4. Press the rotary Adjust/Enter control to commit the change.
The display will return to the normal display.
M370078-01