Non-Isolated Resistive Programming Of Current Calibration - Xantrex Watt Series Programmable DC Power Supply XG 850 Operating Manual

Xg 850 watt series programmable dc power supply
Table of Contents

Advertisement

Non-isolated Resistive Programming of Current Calibration

Important:
calibration procedure again but use the intended operating resistance level in
steps 2 and 3.
To calibrate the non-isolated resistive programming of current:
1. Short the main output with a shunt. Connect a multimeter to measure
2. Connect the 4.000 kΩ resistor to the non-isolated connector (APG -
3. Set the resistive APG level to 4 kΩ. The SCPI command is:
4. Set the input source as Resistive Non-Isolated for Current
5. Turn the main output to On by pressing the OUTPUT ENABLE Main
6. Enter the interactive calibration mode by sending the SCPI command:
7. Press the terminal keys – or + to reduce or increase the main output
8. Press any key other than the + or – keys to exit from interactive
M370078-01
For maximum accuracy at a specific APG level. Follow the
the voltage across the shunt.
J1), one terminal to resistive programming of current (J1.12) and
voltage programming of current (J1.10), and the other terminal to the
common (J1.7).
SYST:COMM:APR:LEV:CURR
Programming. The SCPI command is:
SYST:REM:SOUR:CURR
button on the front panel or send the command:
:OUTP ON
The OUTPUT ENABLE Main button will illuminate.
CAL:OUTP:ANAL:CURR:RES
current value. Adjust the main output current to 100%.
calibration mode.
Calibration and Troubleshooting
4
ARES
6
6-15