Calibrate O 2 Sensor - Dräger Fabius OS Instructions For Use Manual

Anesthesia workstation
Hide thumbs Also See for Fabius OS:
Table of Contents

Advertisement

Calibrate O
Sensor
2
To calibrate the oxygen sensor correctly, make sure
it is exposed only to room air during the entire cali-
bration period.
To avoid leakage, remove the oxygen sensor
assembly from the inspiratory valve dome, and
insert the valve dome plug into the dome.
1 Press the »Calibrate O2 Sensor« soft key in
the Setup window.
The calibration instruction window replaces the
Setup window. Follow the directions provided:
– Remove O
Sensor and expose to room air for
2
2 minutes.
– To start O
calibration press rotary knob.
2
– Observe Calibration status in O
– Reinsert O
Sensor after successful Calibra-
2
tion.
Instructions for Use Fabius OS SW 3.n
data window.
2
Configuration
1
129