Operation
1 Press the »ManSpont« key. The LED associa-
ted with this key starts blinking. It remains blin-
king until the selected mode of operation is
confirmed.
2 The Waveform window is replaced by the
ManSpont window and a message that provi-
des instructions to confirm the mode change.
Confirm the mode change. The ManSpont
screen is activated. After the mode change is
confirmed, the »ManSpont« key LED switches
from blinking to constantly on and the waveform
is restored.
The ManSpont screen allows the user to adjust two
parameters: Apnea Pressure alarm ON/OFF and
Volume alarms ON/OFF. Pressing the »ON/OFF«
soft key toggles the corresponding alarm between
ON and OFF.
Press the O
flush button, as required, to rein-
2
flate the bag.
Set the appropriate fresh gas flow.
Start manual ventilation. The pressure will be
limited to the value set on the APL valve.
Start manual ventilation by hand via the brea-
thing bag.
To temporarily relieve pressure:
Pull up on the APL valve knob.
78
1
10
20
Instructions for Use Fabius OS SW 3.n
2
30