Dräger Fabius OS Instructions For Use Manual page 113

Anesthesia workstation
Hide thumbs Also See for Fabius OS:
Table of Contents

Advertisement

Compliance Test Results
The system compliance test determines the current
compliance of the patient system with any filters,
hoses, and Y-piece. The compliance value is used
during Volume Control mode ventilation to correct
for the reduction of tidal volume due to system com-
pliance. A system compliance value of up
6.5 mL/cmH
O (6.5 mL/hPa) is posted as
2
»PASSED« on the leak test result screen, and the
compliance value is also displayed on the Standby
screen.
Ventilator Leak Test Results
If a ventilator leak test is posted as »PASSED«, the
value is displayed on the leak test result screen and
the Standby screen.
Measured ventilator
leak [mL/min]
≤150
151 to 250
>250
System Leak Test Results
System test results are posted on the leak test
result screen.
Measured system leak
[mL/min]
≤250
251 to 350
>350
Instructions for Use Fabius OS SW 3.n
Displayed result
[mL/min]
measured value and
PASSED
measured value and
FAILED
>250 and FAILED
Displayed result
[mL/min]
measured value and
PASSED
measured value and
FAILED
>350 and FAILED
Fresh-gas
Ventilator
Decoupling
Vaporizer
Flow Tube
O
2+
Oxygen
Flush
Y-piece: closed
Fresh-gas
Ventilator
Decoupling
Vaporizer
Flow Tube
O
2+
Oxygen
Flush
Configuration
Y-piece: closed
Sample line: closed
Setting
P
/
MAX
PEEP
APL
APL
Bypass
Valve
Absorber
Bag
Scavenging
Sample line: closed
Setting
P
/
MAX
PEEP
APL
APL
Bypass
Valve
Absorber
Bag
Scavenging
113