Form - Dräger Fabius OS Instructions For Use Manual

Anesthesia workstation
Hide thumbs Also See for Fabius OS:
Table of Contents

Advertisement

Appendix – Daily and Pre-use Checkout Form
Before operating the Fabius OS, the following
checkout verification form must be completed to
ensure that the machine is ready for use. Do not
insert any additional components into or modify the
anesthesia system after the checkout procedure is
started.
This is a recommended procedure. Follow your
institution's policies for specific checkout procedu-
res.
CAUTION
If any check cannot be carried out satisfactorily,
the machine must not be used.
Call DrägerService or your local authorized ser-
vice organization.
NOTE:
Throughout this section, cmH
Instructions for Use Fabius OS SW 3.n
O = mbar = hPa
2
Appendix – Daily and Pre-use Checkout Form
195

Advertisement

Table of Contents
loading

Table of Contents