Volume Control Ventilation

Ventilator Compliance Compensation
Ventilator compliance compensation is continu-
ously applied during Volume Control so that the
tidal volume delivered to the patient corresponds to
the V
setting. Ventilator compliance is determined
T
during the leak and compliance test performed from
the Standby mode (see "Leak/Compliance Test" on
page 112). To have compliance compensation work
accurately, it is important that the patient hoses
used during the leak/compliance test match the
type of hoses used during the procedure.
NOTE:
When the ventilator settings for Volume Control
cause the ventilator to operate at its limits of perfor-
mance, it is not possible for the Fabius OS to apply
compliance compensation. If the ventilator's perfor-
mance limit is reached, it is not possible to incre-
ment the V
setting via the Volume Control Settings
T
window.
Starting Volume Control Ventilation
The following examples and illustrations describe
starting Volume Control ventilation from the present
ventilation mode "Pressure Control":
1 Press the »Volume Control« key. The LED
associated with this key starts blinking. It
remains blinking until the selected mode of ope-
ration is confirmed.
2 The Waveform window is replaced by the venti-
lator settings window and a message that provi-
des instructions to confirm the mode change.
Instructions for Use Fabius OS SW 3.n
1
2
Operation
79