Dräger Fabius OS Instructions For Use Manual page 157

Anesthesia workstation
Hide thumbs Also See for Fabius OS:
Table of Contents

Advertisement

Fabius OS
Recommen-
Components
ded
which can be
reprocessing
reprocessed
intervals
Other
AGS scavenging
Daily
system
Suction system and
Daily
suction bottle
assembly
accessories
1) It is generally recommended to perform a pre-cleaning under running water for approx. 5 minutes.
For dried soiling an additional pre-cleaning step using an ultrasonic bath is recommended. The addition of a quaternary
ammonium compound supports the cleaning performance effectively (e. g. Gigasept AF). Perform pre-cleaning in the ultra-
sonic bath for approx. 15 minutes. Afterwards rinse under clear running water for approx. 5 minutes (waste water must be
free of visible cleaning residues).
2) Use mild alkaline or neutral cleaner
3) Use disinfectants based on aldehydes and quaternary ammonium compounds.
4) Do not use any agents containing alcohol.
5) After the machine cleaning and disinfection, a hot steam sterilization is required to dry the breathing system. Insufficient
drying of the control areas found in the valve plate can lead to an adverse effect on the functions of the medical device or
to its failure!
6) Valves, ventilator cover and AGS sleeve: Make sure that the cleaning and rinsing liquids can flow in the opening direction
of the valve.
7) Remove any water which may have accumulated in the ventilator diaphragm. Large quantities of condensed water can
negatively effect the operation of the device and lead to failure of the medical device!
Instructions for Use Fabius OS SW 3.n
Manual pre-
Machine
1)
cleaning
cleaning and
thermal
disinfection
Reprocessing according to appropriate Instructions for Use.
Reprocessing according to appropriate Instructions for Use.
Cleaning, Disinfection and Sterilization
Manual disinfection
Surface
Immersio
disinfection
n bath
2)
disinfec-
tion
3)
Hot steam
sterili-
zation
157