Working With These Instructions For Use - Dräger Fabius OS Instructions For Use Manual

Anesthesia workstation
Hide thumbs Also See for Fabius OS:
Table of Contents

Advertisement

Working with these
Instructions for Use
Header Line
The header line on each page contains the title of
the chapter.
Page Body
The page body in these Instructions for Use combi-
nes text and illustrations. The information is pre-
sented as sequential steps of action, giving the user
hands-on experience in learning how to use the
Fabius OS machine.
Left-Hand Column – the Text
The text in the left-hand column provides explana-
tions and step-by-step instructions on the practical
use of the machine.
Bullet points indicate separate actions.
1 Where several actions are described, numbers
are used to refer to relevant details in the illust-
rations. On each page the numbering restart
with "1".
– Dashes indicate the listing of data, options or
objects.
Right-Hand Column – the Illustrations
The illustrations provide visual reference for the
text and for locating the various parts of the equip-
ment. Elements mentioned in the text are highligh-
ted. Renderings of screen displays guide the user
and provide a way to reconfirm actions performed.
Typing Conventions in this Manual
– User controls, such as hard keys and soft keys,
and screen pages are printed in bold within quo-
tation marks, e.g., »PEEP« or »Volume Set-
tings«.
– Screen messages are printed in bold within
quotation marks, e.g., »Flow Calibration in
progress«.
– Alarm messages are printed in bold within quo-
tation marks, including the exclamation marks
that indicate their alarm urgency level, e.g.,
»APNEA PRESSURE!!!«.
2
Configuration
Configuration Functions in Standby
Mode
The configuration functions available in Standby
include calibrations, system tests, and the manage-
ment of default settings.
To access Standby mode:
1 Press the »
« key (Standby).
The waveform window is replaced by a confirma-
tion message and the instruction to shut off flow.
The LED on the Standby key starts blinking and will
remain blinking until Standby is confirmed.
NOTE:
If confirmation does not occur within 15 seconds,
the ventilator remains in the previous mode and
the waveform window is restored.
Confirm the mode change. The ventilator enters
Standby mode, the Standby screen replaces the
previous screen, and the Standby LED stops
blinking and remains on.
The following soft key labels appear at the bot-
tom of the Standby screen:
– »Run System Test«
– »Calibrate Flow Sensor«
– »Calibrate O
Sensor«
2
– »Leak/Compl Test«
– »Access Alarm Log«
– »Restore Site Defaults«
Turn off fresh gas flow.
118
Instructions for Use Fabius OS SW 3.n
1
Instructions for Use Fabius OS SW 3.n