Dräger Fabius OS Instructions For Use Manual page 200

Anesthesia workstation
Hide thumbs Also See for Fabius OS:
Table of Contents

Advertisement

Appendix – Daily and Pre-use Checkout Form
Additional Monitors (when present)
Check the CO
2
Check the anesthetic agent monitor and
alarm module.
Anesthetic Gas Scavenging System
P
Check the hose connections.
P
Adjust the flow regulator to place the float
between the "Minimum" and "Maximum"
marks.
P
Close all flow control valves on the machine
with Y-piece occluded.
P
Change to Standby Screen.
– Select the APL valve for Spontaneous
Breathing:
Rotate the APL valve knob fully counter-
clockwise to spont mark.
Press and hold the O
verify that airway pressure is less than
10 cmH
O (hPa) with Y-piece occluded.
2
– Release the O
that airway pressure is equal or greather
0 cmH
O (hPa).
2
Manual Ventilation Bag for Emergency Ventila-
tion
Check that the bag is functioning correctly
by pumping manually.
When the bag is squeezed, air must audibly
and tangibly flow out of the mask cone;
when the bag is released, it must rapidly
recover its original shape.
Block off the mask connector (cone) with the
ball of your thumb: you should only be able
to squeeze the bag a little.
200
monitor and alarm module.
flush button and
2
Flush button and verify
2
P
Before Connecting to Patient
Verify that
– all vaporizers are off (the handwheels
are set to zero),
– the APL Valve is set as desired,
– all flowmeters indicate 0,
– the patient suction is level adequate,
and
– the breathing system is ready to use (the
bag is in place and all hoses are connec-
ted properly).
If any check can not be carried out satisfacto-
rily, the machine must not be used.
Daily Checkout Signature
Name
Date
Pre-use Checkout Signature
Name
Date
Pre-use Checkout Signature
Name
Date
Pre-use Checkout Signature
Name
Date
Instructions for Use Fabius OS SW 3.n