Dräger Fabius OS Instructions For Use Manual page 120

Anesthesia workstation
Hide thumbs Also See for Fabius OS:
Table of Contents

Advertisement

Configuration
The adjustment range and factory default values for
all alarms on the Fabius OS are shown in the follo-
wing table.
Alarm
Adjustment
Parameter
Range
O
19 to 100
2
%
18 to 99
MV
0.1 to 20.0
L/min
0.0 to 19.9
Pressure
10 to 70
cmH
O
5 to 30
2
(hPa)
Alarm Volume
Select and confirm the »Alarm Volume« label
on the Standby Setup screen. The current alarm
volume value appears on the screen.
Select and confirm the new alarm volume value
from 1 (minimum) to 10 (maximum)
(range: >45 dB (A) to <85 dB (A)).
In the example illustration, the value was changed
from »4« to »5«.
The Alarm Volume Setting window is then removed
from the screen, and the cursor returns to the return
arrow.
120
Factory
Default Value
100
20
12.0
3.0
40
8
Instructions for Use Fabius OS SW 3.n