Advertisement

Preparing for Storage or Transport

WARNING
When moving the anesthesia work-
station, remove all monitors and
equipment from the top shelf and
hinged arms, remove the absorber
system, vaporizers, and reserve gas
cylinders, push in the writing tray.
The anesthesia workstation should only be
moved by people who are physically capable
of handling its weight. Dräger recommends
that two people move the anesthesia worksta-
tion to aid in maneuverability.
Exercise special care so that the machine
does not tip when moving up or down
inclines, around corners, and across thresh-
olds (for example, in door frames and eleva-
tors). Do not attempt to pull the machine over
any hoses, cords, or other obstacles on the
floor.
1 Press and confirm the »
Monitoring and alarms are turned off and the venti-
lator stops.
Turn vaporizers off by turning the handwheel to
»0« until the button engages.
Turn off fresh gas flow.
Close the cylinder valves.
Remove the O
sensor from the inspiratory
2
valve and leave it exposed to air. This precau-
tion prolongs the service life of the sensor.
2 Switch off system power using the switch on the
back of the machine, and disconnect the power
plug.
Disconnect the scavenger hoses.
Instructions for Use Fabius OS SW 3.n
« key (Standby).
Operation
1
2
91