O 2 Sensor - Dräger Fabius OS Instructions For Use Manual

Anesthesia workstation
Hide thumbs Also See for Fabius OS:
Table of Contents

Advertisement

O
Sensor
2
CAUTION
Risk of equipment damage
The O
Sensor must not be sterilized or disinfec-
2
ted!
Spirolog Infinity ID and SpiroLife flow sensors
Reprocess the flow sensor in accordance with the
corresponding Instructions for Use.
CAUTION
The flow sensor must not be reprocessed in a
washer-disinfector. Do not clean with compressed
air, water jet, brush, etc. Otherwise, the thin wires
in the flow sensor may be destroyed. The Spirolog
and Infinity ID flow sensors must not be sterilized
in hot steam. The flow sensors are not resistant to
high temperatures and will be destroyed.
CAUTION
Only use clean disinfectant solutions to disinfect
the flow sensor. Contaminants, e.g., lint, may lead
to the destruction of the flow sensor.
WARNING
Risk of fire hazard
Allow the flow sensor to dry in air for at least
30 minutes after the use of disinfectants
containing flammable substances. These
substances give off vapors that could ignite
during calibration.
CAUTION
The flow sensor can only be reused as long as
automatic calibration is possible.
Instructions for Use Fabius OS SW 3.n
Cleaning, Disinfection and Sterilization
Sterilization
CAUTION!
Risk of damage to the medical device
The Spirolog and Infinity ID flow sensors must not
be sterilized!
Strictly follow the correct Instructions for Use.
Steam sterilization at 273 °F (134 °C)
CAUTION
Risk of damage to the medical device
SpiroLife flow sensors are not suitable for plasma
or radiation sterilization.
Strictly follow the correct Instructions for Use.
All components with the suitable reprocessing opti-
ons are listed in the maintenance list for Fabius OS
components page 156. Follow the hygiene regula-
tions of the hospital!