After completion of cleaning and disinfection
program
Immediately remove parts from washer-disin-
fector.
Inspect parts for visible soiling and damage. If
necessary, repeat cycle or clean manually.
Allow parts to dry thoroughly.
Cleaning agents
The material compatibility of reusable Dräger
accessories has been tested with various mildly
alkaline and enzymatic cleaning agents at 95 °C
(203 °F) for 10 minutes.
The following cleaning agents showed good mate-
rial compatibility at the time of the test:
– Neodisher FA, Neodisher Medizym manufactu-
red by Dr. Weigert
The user must strictly observe the manufacturer's
instructions for use for the cleaning agent.

Manual cleaning

If no washer-disinfector is available, clean parts
manually under running water with commercially
available cleaning agents. The user must strictly
observe the manufacturer's instructions for use for
the cleaning agent.
Wash off soiling on surfaces under running
water.
Use cleaning agents in accordance with manuf-
acturers specifications. Make sure that all sur-
faces to be cleaned can be efficiently reached
(e.g. inside hoses). Use suitable brushes if
necessary.
Do not use any brushes for the flow sensor.
Observe the relevant Instructions for Use.
Rinse parts under running water until no clea-
ning agent residues are discernible.
Check parts for visible soiling and damage.
Repeat manual cleaning if necessary.
Instructions for Use Fabius OS SW 3.n
Cleaning, Disinfection and Sterilization

Manual disinfection

Manual disinfection should preferably be carried
out with disinfectants based on aldehydes or qua-
ternary ammonia compounds. The efficiency of the
disinfectants used must be proven. Observe the
applicable country-specific listings. The user must
strictly observe the manufacturer's instructions for
use for the cleaning agent.
Disinfectants
The material compatibility of Dräger accessories to
be reprocessed has been tested with various disin-
fectants.
The test showed that the following disinfectants
have a good material compatibility:
Surface disinfectant (for device surfaces)
– Incidin Extra N from Ecolab
– Incidur from Ecolab
Instrument disinfectant (for components or acces-
sories):
– Korsolex extra manufactured by Bode Chemie
– Gigasept FF manufactured by Schülke & Mayr
The composition of the disinfectant is the responsi-
bility of the manufacturer and can change over
time.
Disinfecting surfaces
WARNING
Penetrating liquid may lead to failure of the
medical device or damage to the medical
device and endanger the patient! Only
disinfect parts by wiping and make sure no
liquids penetrate into the device.
Following manual cleaning, carry out surface
disinfection.
Remove disinfectant residues.