  • Page 3 Introduction This kit includes parts to upgrade a N9032B PXA signal analyzer to add option EXM, External Mixing. This kit includes parts to upgrade a PXA regardless of frequency range option installed. In addition to installing the hardware to support option EXM, a license for the following option will also be installed: —...
  • Page 4 Option EXM, External Mixing Upgrade Kit Installation Kit Parts List Quantity Description Keysight Part Number Adapter- Coaxial Straight Female-SMA to Female-SMA, 50Ω 1250-1666 Washer, Lock, Internal Tooth, ¼ inch 2190-0067 Nut-Hex-Double-Chamfer 1/4-36-THD.125-IN-THK, Stainless Steel 2950-0223 Cable Assembly, External Mixing, Front Panel (W30)
  • Page 5 — 5/16-inch torque wrench — ¼-inch open-end wrench — Small flat blade screwdriver — Keysight Calibration and Adjustment Software — Test equipment and computer supported by the N78111B PXA Calibration Application — Microsoft Windows based personal computer with internet access and USB port —...
  • Page 6: Analyzer Disassembly

    Option EXM, External Mixing Upgrade Kit Analyzer Disassembly If the instrument is placed on its face during any of the following procedures, be sure to use a soft surface or soft cloth to avoid damage to the front panel, keys, or input connector. Make sure any adapters on the front panel are removed.
  • Page 7 Option EXM, External Mixing Upgrade Kit Top Brace Removal Refer to Figure 2. To remove the top brace (1), use the T-10 driver to remove the twelve screws (2) (0515-0372) attaching the top brace to the chassis. Remove the wire hold down (as shown in Figure 3).
  • Page 8 Option EXM, External Mixing Upgrade Kit Figure 3 Wire Hold Down Installation Note N9032-90004...
  • Page 9: Front Frame Assembly Removal

    Option EXM, External Mixing Upgrade Kit Front Frame Assembly Removal Make sure any connectors on the front panel are removed. 1. Remove the plastic side trim strips by inserting a small flat blade screwdriver between the bottom of the trim strip and the frame. Then carefully working up the side of the trim strip to remove it.
  • Page 10 Option EXM, External Mixing Upgrade Kit Figure 5 Front Panel Cable Installation Note N9032-90004...
  • Page 11 Option EXM, External Mixing Upgrade Kit Add a Cable to the A13 Front End Assembly J8 1. Remove the SMA termination on A13A1J8. A13A1J8 is to the left of A13A1J13. 2. Save this SMA termination. It will be installed on the front panel EXT mixer connector. 3.
  • Page 12: Add Ext Mixer Connector And Cable To Front Frame Assembly

    Option EXM, External Mixing Upgrade Kit Add EXT MIXER Connector and Cable to Front Frame Assembly 1. Locate the SMA female to SMA female connector (1250-1666), ¼” lockwasher (2190-0067), and ¼” hex nut (2950-0223) in the kit. If the SMA connector includes hardware, discard that hardware and use the 2190-0067 lockwasher and...
  • Page 13 Option EXM, External Mixing Upgrade Kit Figure 8 Connecting W30 to Ext Mixer Connector Figure 9 Proper Orientation of W30 Installation Note N9032-90004...
  • Page 14 Option EXM, External Mixing Upgrade Kit Front Frame Replacement 1. Refer to Figure 10. Reattach the ribbon cable W1. Ensure the locking tabs are engaged. 2. Refer to Figure 11. Carefully position the Front Frame Assembly onto the chassis. Ensure no cables are crushed.
  • Page 15 Option EXM, External Mixing Upgrade Kit Figure 11 Front Frame Replacement Installation Note N9032-90004...
  • Page 16 Option EXM, External Mixing Upgrade Kit 3. Refer to Figure 12. When the front panel is installed, cables W29 and W30 typically do not align perfectly. This is OK since the cables are slightly flexible. If additional alignment is needed loosen W29 at A13J8 and re-position. 4.
  • Page 17 Option EXM, External Mixing Upgrade Kit Figure 13 W29 and W30 Properly Aligned 5. Refer to Figure 14. Use a ¼” open-end wrench to prevent the SMA female connector on W29 from twisting. Use a 5/16” torque wrench to torque the nut on W30 to 10 inch-pounds. Figure 14 Torque Cable W30 onto W29 Installation Note N9032-90004...
  • Page 18 Option EXM, External Mixing Upgrade Kit Top Brace Replacement 1. Refer to Figure 15. To replace the top brace, place the brace in the correct position and attach to the chassis using the twelve screws (2) (0515-0372). Attach the top brace using the seventeen screws (3) (0515-1946) included in the kit.
  • Page 19 Option EXM, External Mixing Upgrade Kit Figure 16 Wire Hold Downs 4. In the upgrade kit, locate the Warning Label, N9030-80018. 5. Refer to Figure 17. Attach the Warning Label to the top brace as shown. Figure 17 Add Warning Label to Top Brace Installation Note N9032-90004...
  • Page 20: Final Assembly

    Option EXM, External Mixing Upgrade Kit Final Assembly 1. Refer to Figure 18. Slide the instrument cover back onto the deck from the rear. The seam on the cover should be on the bottom. Be sure the cover seats into the gasket groove in the Front Frame Assembly.
  • Page 21 USB Storage Device Configuration Menu 7. The analyzer will automatically consume the License File (this may take a few minutes). When the License File is consumed the Keysight License Manager will display a “Successful License Installation” message similar to the one shown in...
  • Page 22: Verify The License Installation

    Option EXM, External Mixing Upgrade Kit Figure 20 Successful License Installation Alternate Installation Procedure The License File can be manually installed over USB or LAN by placing the license file in the following analyzer folder: C:\Program Files\Agilent\licensing. Verify the License Installation 1.
  • Page 23 For assistance, contact your nearest Keysight Technologies Sales and Service Office. To find your local Keysight office access the following URL, or if in the United States, call the following telephone number: http://www.keysight.com/find/assist...
  • Page 24 This information is subject to change without notice. © Keysight Technologies 2021 Edition 2, October 2021 N9032-90004 www.keysight.com...