Ci Base Calibration - JUMO AQUIS touch P Installation Instructions Manual

Modular multichannel measuring device for liquid analysis with integrated controller and paperless recorder
Hide thumbs Also See for AQUIS touch P:
Table of Contents

Advertisement

10 Retrofitting optional boards
10.2

Ci base calibration

10
Analysis inputs for inductive conductivity sensors must undergo a Ci base cali-
bration in the course of commissioning. A basic Ci calibration must be performed
when:
A new sensor or a new Ci optional board is being installed for the first time
The sensor or a Ci optional board was replaced
A Ci optional board was repositioned to a different optional board slot
Data loss occurred as the result of failure of the backup battery while the
power was switched off
Device software was updated
After the basic calibration has been performed, the measuring input can be cal-
ibrated. Following successful calibration, the measuring input is ready for use.
NOTE!
For the Ci base calibration, you need the JUMO Type 202711/21 calibration
adapter for inductive conductivity sensors (TN 00543395).
CAUTION!
A buffer battery is built into the device. Its purpose is to maintain data when the
device is switched off or if the voltage supply is interrupted. If the battery ap-
proaches the end of its operating life (approximately 7 years) then a battery
prealarm will indicate that status. When the battery is empty a battery alarm will
indicate that status. The battery must be exchanged before it is empty. The bat-
tery must be exchanged by the JUMO Service department! In this case, send
in the device!