18 Technical data
18.2.3

Analysis input: CR (resistive conductivity)

Units
a
Display ranges
Temperature compensation
Cell constant
Measuring range selection
Measuring accuracy
Ambient temperature influence 0.2%/10 K
a
The display range is scalable. The Comma format is freely configurable. In addition, an
automatic Comma format can be set.
b
TDS (Total Dissolved Solids)
c
In the Configuration up to 4 different measuring ranges with separate
display range limits, units,temperature compensation methods, and alarm functions
can be configured. The respective active measuring range is selected via binary signals.
d
MR: measuring range span
66
µs/cm
ms/cm
kΩ × cm
MΩ × cm
0.0000 to 9.9999
00.000 to 99.999
000.00 to 999.99
0000.0 to 9999.9
00000 to 99999
TC-linear,
natural water DIN EN 27888,
natural water with expanded range,
b
TDS
,
ASTM D-1125-95 for neutral (NaCl), acidic (HCl) and alkaline
(NaOH) impurities
-1
0.01 to 10 cm
c
4 configurable measuring ranges
d
≤ 0.6% of MR
+0.3 µS × cell constant (C)