15 Calibrating CR conductivity sensors
15.3
CR calibration routines
15.3.1
Calibrating the relative cell constant
38
NOTE!
Conductivity measuring inputs can be configured with measuring range change-
over. Accordingly, calibrations must be performed for all "accessible measuring
ranges".
NOTE!
You must be logged in with corresponding user rights to perform calibrations.
v Chapter 8.2.1 "Log-on/Log-out", Page 74
Step
Action
1
Start calibration of the relative cell constant.
Device menu r Calibration r Select CR Analysis Input r Rel. cell
constant Calibration
For CR optional boards, continue with step 2; for universal inputs
set to the "conductivity measurement" operation mode, continue
with step 3
2
Enter one of the measuring ranges 1 to 4. Confirm the entry by
pressing "OK"
The calibration values determined apply only to the selected mea-
suring range.
Tap button to enter
measuring range