Calibration default settings for calibrating the relative cell constant
Parameter
Reference conductivi-
ty
Calibration default settings for calibrating the temperature coefficient
Parameter
Temperature com-
pensation
Reference tempera-
ture
Operation tempera-
ture
15 Calibrating CR conductivity sensors
Setting options
0 to 9999 mS/cm
Setting options
Selection from
analog selection
-50 to +150° C
-50 to +150° C
Explanation
Conductivity of the reference solution
Explanation
Temperature input for automatic sensing of the test/
sample solution temperature during the calibration
The conductivities of a sample solution at the refer-
ence temperature and operation temperature are
sensed during the calibration procedure. This yields
2 value pairs (temperature/conductivity). These val-
ue pairs provide the basis for calculating the tem-
perature coefficient.
The operation temp. must differ from the reference
temperature by at least 5 °C.
37