13 Calibrating Redox sensors
13.2.2

Calibration default settings for Redox sensors

Redox calibration default settings
Parameter
Redox test solution
30
Before you can perform a calibration, you must first enter the necessary calibra-
tion default settings. The possible settings for the Redox calibration are de-
scribed in the following.
Open the calibration default settings:
Device Menu r Calibration r Select Analysis Input for pH/Redox/
r Calibration Default Settings
NH
3
NOTE!
The "Calibration Default Settings" menu appears in the Device Menu only if a
user with corresponding user rights is logged in.
v Chapter 8.2.1 "Log-on/Log-out", Page 74
Sample screen:
Redox calibration default
settings
(zero-point calibration)
The calibration default settings enable the calibration routines to be accessed in
the particular calibration menu.
Calibration routines that are not enabled are not visible in the calibration menu.
Additional calibration default settings are explained in the following table.
Possible settings
-1500 to +1500 mV
NOTE!
Remember that the configuration of the Redox measuring input must be set to
the Redox unit "mV" for the zero-point calibration and to "Percent" for the two-
point calibration.
v Chapter 9.5.3 "pH/Redox/NH3 analysis inputs", Page 102
Explanation
Manual entry of the Redox potential of
the test solution being used for calibra-
tion