Calibrating Ammonia Sensors; Notes; General Information; Calibration Methods For Ammonia Sensors - JUMO AQUIS touch P Installation Instructions Manual

Modular multichannel measuring device for liquid analysis with integrated controller and paperless recorder
Hide thumbs Also See for AQUIS touch P:
Table of Contents

Advertisement

14.1

Notes

14.2

General information

14.2.1

Calibration methods for ammonia sensors

Zero-point calibration
14.2.2

Calibration default settings for ammonia sensors

WARNING!
During the calibration, the relays and analog output signals assume the config-
ured states!
The calibration of ammonia sensors is based on measurements in ammonia-
free test solutions.
This calibration method is used to determine the ammonia zero point.
An ammonia-free test solution (e. g. water) is needed as a reference.
In the ammonia sensor calibration default settings, the zero-point calibration is
enabled and preconfigured as the only available calibration routine.
Open the calibration default settings:
Device menu r Calibration r Select Analysis Input for pH/Redox/
r Calibration Default Settings
NH
3
Sample screen:
Ammonia calibration
default settings

14 Calibrating ammonia sensors

33