JUMO AQUIS touch P Installation Instructions Manual page 14

Modular multichannel measuring device for liquid analysis with integrated controller and paperless recorder
Hide thumbs Also See for AQUIS touch P:
Table of Contents

Advertisement

10 Retrofitting optional boards
14
Step
Action
11
Once all measurings have been confirmed, a summary of the cali-
bration data acquired appears. Confirm by pressing "OK".
If the Ci base calibration fails, the procedure is canceled without
acceptance of the calibration data.
Protocol after
successful
Ci base calibration
Protocol after failed
Ci base calibration
12
Accept the calibration data acquired by pressing "Yes"; discard the
data by pressing "No".